Using incompatible devices damages endoscopes Certain types of non-Olympus flexible accessories could damage or even perforate the biopsy channel of your Olympus endoscopes, resulting in costly repairs - so it makes good economic sense to use only authentic Olympus EndoTherapy devices with Olympus flexible endoscopes. |
 | Using kinked, repaired or inappropriate accessories may save money now, but you could end up spending thousands of pounds on repairs in the future.
 |
Why? The channel size of your endoscope is measured in fractions of a millimetre - you accessories must therefore be exactly the right size to pass through safely. For example, Olympus forceps have an extra-flexible section at the distal end, which is designed to match the scope’s maximum angulation. Only the original scope manufacturer knows and understands this requirement. | Olympus accessories should pass smoothly through your scope, however, using kinked or buckled accessories will place serious strain on the internal components of your instruement - DO NOT, therefore, use kinked accessories through your scopes.
Colour Coding Olympus EndoTherapy devices have colour-coded handles, matching the colour indicators on the channel openings of Olympus endoscopes. This colour coding is a vital reference to the endoscope’s channel diameter, so by matching an accessory’s handle colour to that of the channel opening, the correct accessory can be quickly and easily identified, giving instant recognition of the correct scope/accessory combination.Accessories of the same colour code can differ in length, so always verify that the chosen model is long enough for the scope. |
| Handle colour |
Min channel dia |
| Nomenclature Each part of an accessory’s model reference indicates and important feature, as follows:Example: FB - 25 K R - 1 (1) (2) (3)(4) (5)
| (1) Accessory Type: |
| FB Biopsy Forceps |
FG Grasping Forceps |
FD Hot Biopsy Forceps |
| FS Surgical Scissors |
NM Injection Needle |
HX Clip Fixing/Ligating Device |
| NA Aspiration Needle |
CD Coagulation Electrode |
PBD Biliary Drainage Tube |
| PR Cannula |
KD Sphincterotomy Knife |
BML Mechanical Lithotriptor |
| G Guidewire |
SD Diathermy Snare |
BA/BB/BC/BE/BP Balloon Dilator |
| B Balloon Catheter |
PW Washing Pipe |
BW Cleaning Brush |
(2) Model Type: These digits indicate the model within the range and generally the order in which they were developed. |
|
(4) Rotatability: An ’R’ in this position indicates an accessory where the operating mechanism can be rotated independently from the handle. |
|
(5) Reprocessing: Accessories with a ’-1’ at the end of their reference are part of the Olympus ’Refined Reprocessing’ range. |
(3) Length: The final letter(s) in the EndoTherapy model reference indicate(s) the accessory working length. As a general guide, these are the most common: |
|
C K/L Q U SX |
Bronchoscope accessory Gastroscope accessory ERCP accessory Colonoscope accessory Cystoscope accessory |
| In full detail, the working lengths indicated by each of the letter references are: |
| A = 850mm |
B = 960mm |
C = 1050mm |
D = 1150mm |
E = 1200mm |
| F = 1250mm |
G = 1300mm |
H = 1400mm |
J = 1450mm |
K = 1550mm |
| L = 1650mm |
M = 1700mm |
N = 1800mm |
P = 1900mm |
Q = 1950mm |
| R = 2050mm |
S = 2150mm |
T = 2200mm |
U = 2300mm |
V = 2400mm |
| W = 2500mm |
X = 2700mm |
Y = 2900mm |
Z = 3000mm |
LB = 4000mm |
| LD = 5000mm |
ST = 500mm |
SW = 650mm |
SX = 700mm |
| |
| |
 |
For further details on products available from KeyMed, please call our Customer Services Team on (01702) 616333 or complete the on-line enquiry form |
|